Performance of X-Ray Imaging Systems Applied to Intravenous Angiography
Author :
Riederer, S.J.
Author_Institution :
General Electric Co., Medical Systems, P. O. Box 414, NB-924, Milwaukee, WI 53201
Volume :
1
Issue :
1
fYear :
1982
fDate :
7/1/1982 12:00:00 AM
Firstpage :
48
Lastpage :
62
Abstract :
A review is made of the performance of X-ray systems applied to the imaging of dilute amounts of iodine using subtraction methods. Relevant performance criteria are defined and desired performance levels are derived. For a general purpose system these are iodine sensitivity of 1 mg/cm2, cancellation of 10 g/cm2 soft tissue and 2 g/cm2 bone to be equivalent to iodine sensitivity, capability of imaging through 25 cm soft tissue, spatial resolution exceeding 1 lp/mm and acceptable imaging time depending on the application. Conceptual aspects of several subtraction techniques are discussed including temporal, generalized temporal, dual-energy K-edge, dual-energy non-K-edge, three-energy K-edge, and hybrid subtraction. Experimental performance of several representative area and linear detector systems is tabulated according to subtraction algorithm. A table of references is provided.
Keywords :
Angiography; Biological tissues; Biomedical imaging; Bones; Image resolution; Object detection; Optical imaging; X-ray detection; X-ray detectors; X-ray imaging;