DocumentCode :
1143081
Title :
Super PETT I: A Positron Emission Tomograph Utilizing Photon Time-of-Flight Information
Author :
Ter-Pogossian, Michel M. ; Ficke, David C. ; Yamamoto, Mikio ; Hood, John T., Sr.
Volume :
1
Issue :
3
fYear :
1982
Firstpage :
179
Lastpage :
187
Abstract :
The physical characteristics and some imaging capabilities of Super PETT I, a positron emission tomograph utilizing time-of-flight (TOF) in its image reconstruction process were assessed experimentally by means of measurements carried out in phantoms and clinical imaging studies. The performance characteristics assessed included sensitivity, spatial resolution, image improvements resulting from time-of-flight information utilization, system dead time, and linearity. The clinical examples included imaging of the brain, the heart, the liver, and a demonstration of Super PETT I´s capability of achieving cardiac gating.
Keywords :
Biomedical imaging; Gain measurement; Image reconstruction; Positron emission tomography; Radioactive decay; Signal to noise ratio; Single photon emission computed tomography; Spatial resolution; Strontium; Testing;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.1982.4307570
Filename :
4307570
Link To Document :
بازگشت