DocumentCode :
1145097
Title :
Implementing and Accelerating the EM Algorithm for Positron Emission Tomography
Author :
Kaufman, Linda
Volume :
6
Issue :
1
fYear :
1987
fDate :
3/1/1987 12:00:00 AM
Firstpage :
37
Lastpage :
51
Abstract :
Since the publication of Shepp and Vadi´s [ 14] maximum likelihood reconstruction algorithm for emission tomography (ET), many medical research centers engaged in ET have made an effort to change their reconstruction algorithms to this new approach. Some have succeeded, while others claim they could not adopt this new approach primarily because of limited computing power. In this paper, we discuss techniques for reducing the computational requirements of the reconstruction algorithm. Specifically, the paper discusses the data structures one might use and ways of taking advantage of the geometry of the physical system. The paper also treats some of the numerical aspects of the EM (expectation maximization) algorithm, and ways of speeding up the numerical algorithm using some of the traditional techniques of numerical analysis.
Keywords :
Acceleration; Biochemistry; Blood flow; Data structures; Detectors; Geometry; Maximum likelihood detection; Numerical analysis; Positron emission tomography; Reconstruction algorithms;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.1987.4307796
Filename :
4307796
Link To Document :
بازگشت