DocumentCode :
1190260
Title :
A Self-Balancing Bridge with Range and Precision Adequate for a Platinum Thermometer
Author :
Williams, A.J., Jr. ; Mergner, G.C.
Volume :
15
Issue :
3
fYear :
1966
Firstpage :
121
Lastpage :
131
Abstract :
Calorimetry at NBS needs automation of bridge readout and bridge balancing. Bridge-balancing logic allows it to track the thermometer with progressively decreasing deviations as thermometer temperature approaches equilibrium. A step in the next decade does not interrupt since carry is automatic. Thermometer lead resistance is compensated as lead unbalance changes. Readout, at appropriate times, is into punched cards. With thermometer current only 0.002 ampere (to limit heating), the unbalance detector must function with only 0.02 microvolt available for the minimum step (10 microhms), in spite of much larger disturbing voltages, notably, thermal agitation, thermal EMF, and power frequency pickup.
Keywords :
Automation; Bridge circuits; Calorimetry; Detectors; Heating; Logic; NIST; Platinum; Temperature; Thermal resistance;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1966.4313521