DocumentCode :
1191197
Title :
Calibration of Thermal Transfer Standards of RF Voltage
Author :
Clark, Richard F. ; Jurkus, Algirdas P.
Volume :
16
Issue :
3
fYear :
1967
Firstpage :
232
Lastpage :
237
Abstract :
A system of RF voltage calibration in the VHF and UHF ranges is described. The primary voltage standard is a calorimetric power meter whose impedance is determined accurately. The working standards are thermal transfer instruments (thermal converters and micropotentiometers). The methods used to calibrate the working standards against the primary standard are described in detail. Then the calibration of transfer standards submitted by other laboratories is outlined, and procedures and practices which have been found successful are recommended for use by standardizing laboratories.
Keywords :
Calibration; Coaxial components; Impedance measurement; Laboratories; Measurement standards; Power measurement; Power transmission lines; Radio frequency; Transmission line measurements; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1967.4313627
Filename :
4313627
Link To Document :
بازگشت