Title
Calculations of the sensitivity function for a mis-aligned thin film in a Mallinson coil set
Author
Hsu, Shau-Wei
Author_Institution
Center for Meas. Standards, Ind. Technol. Res. Inst., Hsinchu, Taiwan
Volume
30
Issue
6
fYear
1994
fDate
11/1/1994 12:00:00 AM
Firstpage
4584
Lastpage
4586
Abstract
Studies the accuracy of angular calibrations for VSM measurements on a thin film sample in the presence of misalignments have been performed by computations of the sensitivity function. The pick-up coils are the usual Mallison-type. For a square thin-film sample with dimensions of 4 mm×4 mm, the errors due to displacements or off-center rotations of the thin film sample are lower than 5% for all rotation angles if the mis-alignments are controlled to within 1 mm. They increase rapidly, however, if the positional misalignments are larger than 1 mm. If the alignment is performed carefully, the errors caused by tilting the thin film sample is less than 1%. Constraints on the edge length of the thin film is also useful in preventing excessive errors
Keywords
angular measurement; calibration; magnetic moments; magnetic variables measurement; sensitivity; 1 mm; 4 mm; Mallinson coil set; VSM measurements; angular calibration; edge length; errors; errors prevention; misaligned thin film; off-center rotations; pick-up coils; positional misalignments; sensitivity function; square thin-film sample; Anisotropic magnetoresistance; Calibration; Coils; Error correction; Magnetic fields; Magnetic films; Magnetic moments; Measurement standards; Performance evaluation; Transistors;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/20.334156
Filename
334156
Link To Document