Title :
Comparison between the 127I2 Stabilized He-Ne Lasers at 633 nm and 612 nm of the BIPM and the IMGC
Author :
Bertinetto, F. ; Cordiale, P. ; Picotto, G.B. ; Chartier, Jean-Marie ; Felder, R. ; Glaser, M.G.
fDate :
3/1/1983 12:00:00 AM
Abstract :
In November 1981, a comparison between BIPM and IMGC He-Ne lasers stabilized to saturated absorption lines of 127I2, at the wavelengths ¿ = 633 nm and ¿ = 612 nm has been made at the BIPM. At 633 nm, a repeatability and a reproducibility within +1 à 10-11 · ¿ have been obtained. At 612 nm, a repeatability within a few 10-11 · ¿ only for one laser and a reproducibility of ± 2 à 10-10 · ¿ have been obtained.
Keywords :
Absorption; Argon; Curve fitting; Frequency; Laboratories; Laser stability; Least squares methods; Power lasers; Reproducibility of results; Testing;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.1983.4315012