• DocumentCode
    1204183
  • Title

    Test results of an apparatus for calorimetric measurement of AC losses in superconductors

  • Author

    Eikelboom, J.A.

  • Author_Institution
    Netherlands Energy Res. Found. ECN, Petten, Netherlands
  • Volume
    28
  • Issue
    1
  • fYear
    1992
  • fDate
    1/1/1992 12:00:00 AM
  • Firstpage
    817
  • Lastpage
    821
  • Abstract
    A description is given of a novel method to measure calorimetrically AC losses in superconducting samples. The boil-off gas is stored below the liquid helium level in a measuring tube. Losses in the range of 1 mW up to 100 mW can be measured with an accuracy of 3%. Results of measurements of AC losses in superconducting wires are presented. In contrast to measurements with gas-flow meters, there is a negligible background signal and an excellent reproducibility as all measurements are performed at the temperature of boiling helium
  • Keywords
    calorimetry; loss measurement; superconducting cables; AC losses; background signal; boil-off gas; calorimetric measurement; reproducibility; superconducting wires; Calibration; Energy measurement; Fluid flow; Glass; Helium; Loss measurement; Superconducting filaments and wires; Superconducting magnets; Superconductivity; Testing;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.120003
  • Filename
    120003