Design of RF needle applicators for optimum SAR distributions in irregularly shaped tumors
Author :
Zhu, Xi-Li ; Gandhi, Om P.
Author_Institution :
Dept. of Electr. Eng., Utah Univ., Salt Lake City, UT, USA
Volume :
35
Issue :
5
fYear :
1988
fDate :
5/1/1988 12:00:00 AM
Firstpage :
382
Lastpage :
388
Abstract :
The impedance method has been used to obtain specific absorption rate (SAR) distributions for a model of a brain tumor for which the shape was determined from CT head scans. Different locations of 4, 5, or 6 needles were used to obtain optimum arrangements that gave minimum standard deviations (SDs) for the SARs in the volume of the tumor. Further work, presently underway, will develop the thermal model to obtain 3-D temperature distributions for the tumor and its surrounding volume to help design applicators giving tumor temperatures with minimum SDs. The procedure described should be applicable for treatment planning of thermotherapy for a number of tumors for which local hyperthermia by RF needle applicators has been found useful.
Keywords :
biomedical equipment; biothermics; radiation therapy; radiofrequency heating; 3D temperature distribution; RF needle applicators; brain tumor model; head CT scans; impedance method; irregularly shaped tumors; optimum specific absorption rate distribution; thermal model; thermotherapy treatment planning; Applicators; Brain modeling; Impedance measurement; Medical treatment; Needles; Neoplasms; Radio frequency; Shape; Specific absorption rate; Temperature distribution; Brain Neoplasms; Computer Simulation; Humans; Hyperthermia, Induced; Models, Theoretical; Needles; Radio Waves; Tomography, X-Ray Computed;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.1397
Filename :
1397
Link To Document :
بازگشت