DocumentCode
1296904
Title
A recirculating cryogenic hydrogen maser
Author
Hürlimann, Martin D. ; Berlinsky, A. John ; Cline, Richard W. ; Hardy, Walter N.
Author_Institution
Department of Physics, University of British Columbia, Vancouver, B.C., Canada V6T 2A6
Issue
2
fYear
1987
fDate
6/1/1987 12:00:00 AM
Firstpage
584
Lastpage
587
Abstract
This paper reports on the design, theory, and initial testing of a new cryogenic hydrogen maser in which hydrogen (H) atoms circulated back and forth between a microwave-pumped state-selecting source and a small split-ring maser cavity. Initial tests observed no measurable fluctuations in the maser frequency at the level of 6 × 10−13 for measuring times between τ = 1 s and 300 s.
Keywords
Atomic clocks; Atomic measurements; Frequency measurement; Hydrogen; Masers; Temperature measurement; Thermal stability;