DocumentCode
1306017
Title
Design procedure for permanent magnet assemblies with uniform magnetic fields for MRI devices
Author
Podol´skii, A.
Author_Institution
11 Krupskoy St., St. Petersburg, Russia
Volume
36
Issue
2
fYear
2000
fDate
3/1/2000 12:00:00 AM
Firstpage
484
Lastpage
490
Abstract
This paper shows that for magnetic resonance imaging (MRI) permanent magnet assemblies with the pronounced pole pieces, the investigation of magnetic field generated by one pole piece is a way to obtain an interrelation between pole piece diameter, magnet assembly air-gap length, and first and second shim dimensions. The investigation makes it possible to design optimum magnet assemblies
Keywords
magnetic fields; magnetic resonance imaging; permanent magnets; MRI devices; design procedure; first shim dimensions; magnet assembly air-gap length; magnetic resonance imaging; optimum magnet assemblies; permanent magnet assemblies; pole piece diameter; pole pieces; second shim dimension; uniform magnetic fields; Air gaps; Assembly; Helium; Magnetic devices; Magnetic fields; Magnetic resonance imaging; Permanent magnets; Saturation magnetization; Soft magnetic materials; Superconducting magnets;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/20.825819
Filename
825819