Title :
Dimensional correction of high dielectric and magnetic constants determined by S parameters measurements
Author :
Henaux, J.C. ; Crozat, P.
Author_Institution :
Inst. d´Electron. Fondamentale, Univ. Paris Sud, Orsay, France
fDate :
7/19/1990 12:00:00 AM
Abstract :
The measurement of high dielectric constants materials in a coaxial line, using an automatic network analyser (ANA), must take the actual diameter of the sample into consideration. The error on the measured epsilon r value may be approximated by the fractional diameter deviation times epsilon r2. A correction method is proposed.
Keywords :
S-parameters; coaxial cables; magnetic permeability measurement; materials testing; measurement errors; microwave measurement; permittivity measurement; ANA; S parameters measurements; automatic network analyser; coaxial line; correction method; dimensional correction; fractional diameter deviation; geometrical correction; high permeability materials; high permittivity materials; measurement of high dielectric constants materials;
Journal_Title :
Electronics Letters
DOI :
10.1049/el:19900745