• DocumentCode
    1345321
  • Title

    Improving the energy resolution of high-pressure Xe cylindrical ionization chambers

  • Author

    Bolotnikov, Aleksey ; Ramsey, Brian

  • Author_Institution
    NASA Marshall Space Flight Center, Huntsville, AL, USA
  • Volume
    44
  • Issue
    3
  • fYear
    1997
  • fDate
    6/1/1997 12:00:00 AM
  • Firstpage
    1006
  • Lastpage
    1010
  • Abstract
    Results are presented on the development of high-pressure (>50 atm) Xe cylindrical ionization chambers for hard X-ray and low energy gamma-ray astronomy. Two methods to improve the energy resolution of these devices have been tested. One uses a shielding mesh and the other employs no mesh but substitutes a rise-time correction. Both techniques can give similar resolutions, with 2.2% FWHM being measured at 662 keV and 2.4% at 511 keV
  • Keywords
    X-ray astronomy; X-ray detection; astronomical instruments; gamma-ray astronomy; gamma-ray detection; ionisation chambers; shielding; xenon; 511 keV; 662 keV; Xe; energy resolution improvement; hard X-ray astronomy; high-pressure cylindrical ionization chambers; low energy gamma-ray astronomy; rise-time correction; shielding mesh; Aluminum; Cathodes; Ceramics; Electrodes; Electronic equipment testing; Energy resolution; Flanges; Ionization chambers; Preamplifiers; Spectroscopy;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.603794
  • Filename
    603794