• DocumentCode
    1366971
  • Title

    Reduction of metal streak artifacts in X-ray computed tomography using a transmission maximum a posteriori algorithm

  • Author

    De Man, B. ; Nuyts, J. ; Dupont, P. ; Marcha, G. ; Suetens, P.

  • Author_Institution
    Med. Image Comput., Katholieke Univ., Leuven, Belgium
  • Volume
    47
  • Issue
    3
  • fYear
    2000
  • fDate
    6/1/2000 12:00:00 AM
  • Firstpage
    977
  • Lastpage
    981
  • Abstract
    A maximum a posteriori algorithm for reduction of metal streak artifacts in X-ray computed tomography is presented. The algorithm uses a Markov random field smoothness prior and applies increased sampling in the reconstructed image. Good results are obtained for simulations and phantom measurements: streak artifacts are reduced while small, line-shaped details are preserved
  • Keywords
    Markov processes; computerised tomography; image reconstruction; medical image processing; Markov random field smoothness prior; X-ray computed tomography; increased sampling; medical diagnostic imaging; metal streak artifacts reduction; phantom measurements; reconstructed image; simulations; small line-shaped details; transmission maximum a posteriori algorithm; Biomedical imaging; Computed tomography; Dentistry; Image reconstruction; Image resolution; Imaging phantoms; Iterative algorithms; Iterative methods; Markov random fields; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.856534
  • Filename