Abstract :
We have developed a new phoswich detector consisting of an NE213 liquid scintillator surrounded by an NE115 plastic scintillator to measure high energy neutron spectra accurately in a charged-particle mixed field. Differences in the decay time constants of light outputs from the two scintillators make it possible to separate charged-particle-induced and neutron-induced pulses. Experiments were carried out at the RIKEN and at the NIRS cyclotrons. This detector was found to be able to discriminate neutron and gamma-ray events from external-proton events. This detector will be useful to measure high energy neutrons in space
Keywords :
cosmic ray apparatus; cosmic ray neutrons; liquid scintillation detectors; neutron detection; solid scintillation detectors; NE115 plastic scintillator; NE213 liquid scintillator; astronomy; charged particles; cosmic ray apparatus; decay time constants; equipment; high energy neutron spectra; instrument; light outputs; measurement technique; phoswich detector; Current measurement; Energy measurement; Event detection; Extraterrestrial measurements; Gamma ray detection; Gamma ray detectors; Neutrons; Particle measurements; Space charge; Space vehicles;
Journal_Title :