DocumentCode
140225
Title
Optimising the Windkessel model for cardiac output monitoring during changes in vascular tone
Author
Charlton, Peter ; Smith, Johan ; Camporota, Luigi ; Beale, Richard ; Alastruey, Jordi
Author_Institution
Dept. of Biomed. Eng., King´s Coll. London, London, UK
fYear
2014
fDate
26-30 Aug. 2014
Firstpage
3759
Lastpage
3762
Abstract
Algorithms for estimating cardiac output (CO) from the arterial blood pressure wave have been observed to be inaccurate during changes in vascular tone. Many such algorithms are based on the Windkessel model of the circulation. We investigated the optimal analytical approaches and assumptions that make up each algorithm during changes in vascular tone. Several analytical approaches and assumptions were evaluated on data from 15 critically ill patients by comparison with thermodilution measurements. We found that the most accurate algorithms assumed a constant compliance for the duration of the beat. They produced a percentage error of ±31% by maintaining the compliance and outflow terms in the Windkessel model. For any algorithm, the following assumptions gave highest accuracy: (i) outflow pressure into the microcirculation is zero; (ii) end of systole is identified using the second derivative of pressure. None of the tested algorithms reached the clinically acceptable accuracy of ±30%.
Keywords
cardiology; haemodynamics; patient monitoring; Windkessel model; arterial blood pressure wave; cardiac output monitoring; microcirculation; thermodilution measurements; vascular tone; Accuracy; Algorithm design and analysis; Arterial blood pressure; Calibration; Mathematical model; Monitoring; Resistance;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
Conference_Location
Chicago, IL
ISSN
1557-170X
Type
conf
DOI
10.1109/EMBC.2014.6944441