• DocumentCode
    141019
  • Title

    A mock circulatory loop for designing and evaluating total artificial hearts

  • Author

    Love, Holley C. ; Timms, Daniel L. ; Nestler, Frank ; Frazier, O. Howard ; Cohn, William E.

  • Author_Institution
    Texas Heart Inst. at St. Luke´s Hosp., Houston, TX, USA
  • fYear
    2014
  • fDate
    26-30 Aug. 2014
  • Firstpage
    5667
  • Lastpage
    5670
  • Abstract
    A mock circulatory loop was constructed to facilitate total artificial heart development. The loop includes many novel features such as a pressure-regulated tank to simulate exercise conditions, controllable systemic and pulmonary vascular resistance to create left-right flow imbalances as seen in postural change and breathing, and a left atrial suction valve. Dual HeartMate II pumps and the BiVACOR® rotary total artificial heart were used to generate pressure and flow data characterizing the flow loop.
  • Keywords
    cardiology; haemodynamics; prosthetics; BiVACOR; Dual HeartMate II pumps; breathing; flow loop; left atrial suction valve; left-right flow imbalances; mock circulatory loop; postural change; pressure-regulated tank; pulmonary vascular resistance; rotary total artificial heart; total artificial heart development; Fluids; Heart; Hemodynamics; Pumps; Resistance; Rotors; Valves;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2014.6944913
  • Filename
    6944913