DocumentCode
141019
Title
A mock circulatory loop for designing and evaluating total artificial hearts
Author
Love, Holley C. ; Timms, Daniel L. ; Nestler, Frank ; Frazier, O. Howard ; Cohn, William E.
Author_Institution
Texas Heart Inst. at St. Luke´s Hosp., Houston, TX, USA
fYear
2014
fDate
26-30 Aug. 2014
Firstpage
5667
Lastpage
5670
Abstract
A mock circulatory loop was constructed to facilitate total artificial heart development. The loop includes many novel features such as a pressure-regulated tank to simulate exercise conditions, controllable systemic and pulmonary vascular resistance to create left-right flow imbalances as seen in postural change and breathing, and a left atrial suction valve. Dual HeartMate II pumps and the BiVACOR® rotary total artificial heart were used to generate pressure and flow data characterizing the flow loop.
Keywords
cardiology; haemodynamics; prosthetics; BiVACOR; Dual HeartMate II pumps; breathing; flow loop; left atrial suction valve; left-right flow imbalances; mock circulatory loop; postural change; pressure-regulated tank; pulmonary vascular resistance; rotary total artificial heart; total artificial heart development; Fluids; Heart; Hemodynamics; Pumps; Resistance; Rotors; Valves;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
Conference_Location
Chicago, IL
ISSN
1557-170X
Type
conf
DOI
10.1109/EMBC.2014.6944913
Filename
6944913
Link To Document