Realistic and Efficient Modeling of Radiotracer Heterogeneity in Monte Carlo Simulations of PET Images With Tumors
Author :
Stute, Simon ; Vauclin, Sébastien ; Necib, Hatem ; Grotus, Nicolas ; Tylski, Perrine ; Rehfeld, Niklas S. ; Hapdey, Sébastien ; Buvat, Irène
Author_Institution :
IMNC Lab., Paris 7 & Paris 11 Univ., Orsay, France
Volume :
59
Issue :
1
fYear :
2012
Firstpage :
113
Lastpage :
122
Abstract :
Monte Carlo simulations are extensively used in PET to evaluate the accuracy with which PET images can yield reliable estimates of parameters of interest. For such applications, the simulated images should be as realistic as possible so that conclusions can be extrapolated to clinical PET images. In this work, we describe a method for introducing realistic modeling of radiotracer heterogeneity into Monte Carlo simulations of patient PET scans. The modeling of the complex physiological activity distribution in healthy regions is directly based on real patient PET/CT images, and realistic tumor shapes can be included into these regions. This method represents a competitive alternative to the use of complex anthropomorphic phantoms such as the XCAT, that require a fixed activity per structure. The method is extended to the simulation of serial PET scans with tumor changes, as acquired in the context of therapy monitoring, and this extension is validated using a patient study. Using the proposed method, very realistic patient PET images can be produced for evaluation purposes.In addition, a strategy to efficiently simulate many sets of pathological cases, based on a unique background physiological activity distribution, is described and carefully assessed using a numerical phantom. The background activity is simulated only once, while tumors are simulated separately. The data are then recombined in a specific way so that the final image has the same properties as images produced by simulating pathological and tumor activities at the same time.
Keywords :
Monte Carlo methods; positron emission tomography; radioactive tracers; tumours; Monte Carlo simulation; PET images; XCAT; background activity; complex anthropomorphic phantoms; complex physiological activity distribution; healthy regions; numerical phantom; patient PET scans; patient PET-CT images; radiotracer heterogeneity; serial PET scans; therapy monitoring; tumor shapes; Attenuation; Computational modeling; Image reconstruction; Lungs; Phantoms; Positron emission tomography; Tumors; Monte Carlo (MC) simulation; PET imaging; patient monitoring; realistic simulation;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2011.2177277
Filename :
6123180
Link To Document :
بازگشت