• DocumentCode
    1470063
  • Title

    Shielded open-circuited sample holder for dielectric measurements of solids and liquids

  • Author

    Baker-Jarvis, James ; Janezic, Michael D. ; Jones, Chriss A.

  • Author_Institution
    Div. of Radio Frequency Technol., Nat. Inst. of Stand. & Technol., Boulder, CO, USA
  • Volume
    47
  • Issue
    2
  • fYear
    1998
  • fDate
    4/1/1998 12:00:00 AM
  • Firstpage
    338
  • Lastpage
    344
  • Abstract
    This paper presents a technique for permittivity measurements that can be used for either liquids or solid coaxial samples from near dc to gigahertz frequencies. The method uses both a full-mode model of a coaxial open-circuited termination and a capacitive method for permittivity determination. The method requires the use of an impedance bridge and a network analyzer. Measurements indicate good accuracy for the permittivity over a very wide band of frequencies. For high permittivity materials the sample surfaces must be metallized. An uncertainty analysis is also presented
  • Keywords
    coaxial cables; impedance matching; measurement uncertainty; microwave measurement; permittivity measurement; sample holders; capacitive method; coaxial open-circuited termination; dielectric measurements; full-mode model; impedance bridge; liquid samples; loss factor; microwave measurement; network analyzer; permittivity measurement; shielded open-circuited sample holder; solid coaxial samples; tangential field components; uncertainty analysis; Bridges; Coaxial components; Dielectric measurements; Frequency measurement; Inorganic materials; Liquids; Permittivity measurement; Solids; Surface impedance; Wideband;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.744172
  • Filename
    744172