DocumentCode
1494422
Title
DC SQUID preamplifier for DC-SQUID magnetometer
Author
Koshelets, V.P. ; Matlashov, A.N. ; Serpuchenko, I.L. ; Filippenko, L.V. ; Zhuravlev, Y.E.
Author_Institution
Inst. of Radio Eng. & Electron., Moscow, USSR
Volume
25
Issue
2
fYear
1989
fDate
3/1/1989 12:00:00 AM
Firstpage
1182
Lastpage
1185
Abstract
A DC SQUID (superconducting quantum interference device) magnetometer with DC SQUID preamplifier has been designed and experimentally tested. The total magnetometer noise was reduced to the intrinsic noise level of the input DC SQUID, and magnetometer cryogenic containment was minimized by using the DC SQUID preamplifier. Two identical thin-film DC SQUIDs with 15-μm2-area resistively shunted Nb-AlOx-Nb junctions were utilized. The resulting magnetometer had a coupled energy sensitivity ≅3×10-31 J/Hz in the range from 1 Hz to 30 kHz and about 100-dB dynamic range
Keywords
SQUIDs; aluminium compounds; electron device noise; magnetometers; niobium; 1 Hz to 30 kHz; DC SQUID preamplifier; DC-SQUID magnetometer; Nb-AlOx-Nb; coupled energy sensitivity; intrinsic noise level; magnetometer cryogenic containment; total magnetometer noise; Cryogenics; Interference; Noise level; Noise reduction; Preamplifiers; SQUID magnetometers; Superconducting device noise; Superconducting devices; Superconducting thin films; Testing;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/20.92501
Filename
92501
Link To Document