• DocumentCode
    1497589
  • Title

    Bias Field Inconsistency Correction of Motion-Scattered Multislice MRI for Improved 3D Image Reconstruction

  • Author

    Kim, Kio ; Habas, Piotr A. ; Rajagopalan, Vidya ; Scott, Julia A. ; Corbett-Detig, James M. ; Rousseau, Francois ; Barkovich, A. James ; Glenn, Orit A. ; Studholme, Colin

  • Author_Institution
    Dept. of Radiol. & Biomed. Imaging, Univ. of California, San Francisco, CA, USA
  • Volume
    30
  • Issue
    9
  • fYear
    2011
  • Firstpage
    1704
  • Lastpage
    1712
  • Abstract
    A common solution to clinical MR imaging in the presence of large anatomical motion is to use fast multislice 2D studies to reduce slice acquisition time and provide clinically usable slice data. Recently, techniques have been developed which retrospectively correct large scale 3D motion between individual slices allowing the formation of a geometrically correct 3D volume from the multiple slice stacks. One challenge, however, in the final reconstruction process is the possibility of varying intensity bias in the slice data, typically due to the motion of the anatomy relative to imaging coils. As a result, slices which cover the same region of anatomy at different times may exhibit different sensitivity. This bias field inconsistency can induce artifacts in the final 3D reconstruction that can impact both clinical interpretation of key tissue boundaries and the automated analysis of the data. Here we describe a framework to estimate and correct the bias field inconsistency in each slice collectively across all motion corrupted image slices. Experiments using synthetic and clinical data show that the proposed method reduces intensity variability in tissues and improves the distinction between key tissue types.
  • Keywords
    biological tissues; biomedical MRI; data analysis; image motion analysis; image reconstruction; medical image processing; 3D image reconstruction; automated data analysis; bias field inconsistency correction; clinical MR imaging; clinical interpretation; key tissue boundaries; motion-scattered multislice MRI; slice acquisition time reduction; Bismuth; Coils; Image reconstruction; Imaging; Nonhomogeneous media; Sensitivity; Three dimensional displays; 3D reconstruction; Bias field inhomogeneity; bias field inconsistency; fetal brain; multislice MRI; Algorithms; Artifacts; Bias (Epidemiology); Brain; Female; Humans; Image Enhancement; Image Interpretation, Computer-Assisted; Image Processing, Computer-Assisted; Imaging, Three-Dimensional; Magnetic Resonance Imaging; Multidetector Computed Tomography; Pattern Recognition, Automated; Pregnancy; Prenatal Diagnosis; Sensitivity and Specificity;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2011.2143724
  • Filename
    5752245