DocumentCode
149794
Title
A microstrip resonant biosensor for aqueous glucose detection in microfluidic medical applications
Author
Schwerthoeffer, Udo ; Warter, Christopher ; Weigel, Robert ; Kissinger, Dietmar
Author_Institution
Institute for Electronics Engineering, Cauerstr. 9, 91058 Erlangen, Germany
fYear
2014
fDate
19-23 Jan. 2014
Firstpage
55
Lastpage
57
Abstract
A highly sensitive biosensor to detect small variations of glucose concentrations in water is proposed. The sensor structure is based on a planar half-wavelength (lambda/2) microstrip line (MSL) resonator with a microflu-idic container placed on top. The interaction between the electric field and the milliliter range water-glucose solutions under test can be observed measuring the phase shift of the S-parameter from 3.95 to 4.10 GHz. Combining the microwave detection and the microfluidic network, a time efficient and low cost architecture is presented. With this technique, a very sensitive water-glucose concentration of less than 50 mg/dl can be separated.
Keywords
Containers; Diabetes; Microfluidics; Microstrip; Microwave theory and techniques; Reflection; Sugar; blood sugar; diabetes; glucose; microfluidics; microstrip line res-onator; microwave biosensor;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Wireless Technologies, Networks, and Sensing Systems (BioWireleSS), 2014 IEEE Topical Conference on
Conference_Location
Newport Beach, CA, USA
Print_ISBN
978-1-4799-2298-7
Type
conf
DOI
10.1109/BioWireleSS.2014.6827728
Filename
6827728
Link To Document