    Preamplifier with a second-order high-pass filtering characteristic

  • Author

    Lin, Yue-Der ; Tsai, Chang-Da ; Huang, Hui-Hsun ; Chiou, Dah-Chuan ; Wu, Chien-Ping

  • Author_Institution
    Dept. of Electr. Eng., Nat. Taiwan Univ., Taipei, Taiwan
  • Volume
    46
  • Issue
    5
  • fYear
    1999
  • fDate
    5/1/1999 12:00:00 AM
  • Firstpage
    609
  • Lastpage
    612
  • Abstract
    A new preamplifier for suppressing low-frequency interference is presented. The proposed preamplifier, with its front end being implemented by an instrumentation amplifier, enjoys the following advantages: differential high-pass filtering, high input impedance, high common-mode rejection ratio and low passive sensitivity. This circuit can be realized with commercial operational amplifiers with enough phase margin, or fabricated in a chip for practical measurement of physiological signals.
  • Keywords
    bioelectric potentials; biomedical electronics; high-pass filters; interference suppression; operational amplifiers; preamplifiers; biomedical circuit; chip fabrication; commercial operational amplifiers; differential high-pass filtering; high common-mode rejection ratio; high input impedance; instrumentation amplifier; low passive sensitivity; low-frequency interference suppression; phase margin; physiological signals measurement; second-order high-pass filtering characteristic; Circuits; Differential amplifiers; Filtering; Impedance; Instruments; Interference suppression; Operational amplifiers; Passive filters; Phase measurement; Preamplifiers; Electric Impedance; Equipment Design; Sensitivity and Specificity; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.759062
  • Filename
    759062