• DocumentCode
    1509013
  • Title

    An artefact suppressing fast-recovery myoelectric amplifier

  • Author

    Thorsen, R.

  • Author_Institution
    Dept. of Signal & Systems, Twente Univ., Enschede, Netherlands
  • Volume
    46
  • Issue
    6
  • fYear
    1999
  • fDate
    6/1/1999 12:00:00 AM
  • Firstpage
    764
  • Lastpage
    766
  • Abstract
    An amplifier for recording myoelectric signals using surface electrodes has been developed. The special features are suppression of stimulation artefacts and motion artefacts from electrodes. It is designed for recording of myoelectric signals from a muscle that is being stimulated with short impulses. The artifact suppression is achieved by using fast-recovery instrumentation amplifiers and having a nonlinear feedback loop for automatic compensation of changes in DC-offset.
  • Keywords
    amplifiers; biomedical electrodes; electromyography; medical signal processing; DC-offset changes; artefact suppressing fast-recovery myoelectric amplifier; automatic compensation; muscle stimulation; myoelectric signals recording; nonlinear feedback loop; short impulses; spinal cord lesion people; surface electrodes; Cutoff frequency; Electrodes; Electromyography; Feedback loop; Filters; Muscles; Pulse amplifiers; Resistors; Signal design; Switches; Artifacts; Electromyography; Equipment Design; Feedback; Humans; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.764955
  • Filename
    764955