An improved approximation for the spatial impulse response of a rectangular transducer
Author :
Teo, Tat-Jin
Author_Institution :
Boston Sci. Corp., San Jose, CA, USA
Volume :
45
Issue :
1
fYear :
1998
Firstpage :
76
Lastpage :
83
Abstract :
One common approximation for a field point located in the far field of the transducer is to trapezoidally-shaped function to approximate the true spatial impulse response. A quadrilaterally shaped function is proposed here. The quadrilaterally shaped function also includes the trapezoidally shaped function as a special case. This paper shows that the quadrilateral approximation applied in the specifically defined field location is superior to the trapezoidal approximation for the true spatial impulse response function when the transducer elements have the aspect ratio of approximately 10, typical of those found in 1.5 D arrays.
Keywords :
acoustic field; acoustic transducers; approximation theory; transient response; ultrasonic transducers; 1.5 D array; aspect ratio; far field; quadrilateral approximation; rectangular transducer; spatial impulse response; trapezoidal approximation; Apertures; Biomedical imaging; Biomedical transducers; Convolution; Fourier transforms; Geometry; Sampling methods; Ultrasonic imaging; Ultrasonic transducers;
fLanguage :
English