Title :
A phase-measurement error compensation technique suitable for automation
Author :
Wagdy, Mahmoud Fawzy ; Lucas, M.S.P.
Author_Institution :
Dept. of Electr. & Comput. Eng., Northeastern Univ., Boston, MA, USA
fDate :
3/1/1986 12:00:00 AM
Abstract :
A phase meter is presented which incorporates zero crossing detectors (ZCDs) with hysteresis. The system performance is good when the ZCDs are bipolar comparators, but worse when they are operational amplifiers whose switching times are heavily dependent on both the amplitude and frequency of the input signal. To improve the performance, a new error correction technique is introduced which has been successfully implemented in the laboratory and is suitable for use in automated instruments. The technique is based on measuring the erroneous phase angle, then isolating the error sources and measuring the error angle separately. The difference between the two measured values is the correct phase. Thus using a computer in conjunction with lower quality hardware, the same results as with higher quality hardware were achieved.
Keywords :
error compensation; error correction; measurement errors; phase measurement; phase meters; automation; erroneous phase angle; error angle; error correction; hysteresis; laboratory; phase meter; phase-measurement error compensation; zero crossing detectors;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.1986.6499048