• DocumentCode
    1554822
  • Title

    Uranium pinhole collimators for I-131 SPECT imaging

  • Author

    Tenney, C.R. ; Smith, M.F. ; Greer, K.L. ; Jaszczak, R.J.

  • Author_Institution
    Duke Univ. Med. Center, Durham, NC, USA
  • Volume
    46
  • Issue
    4
  • fYear
    1999
  • fDate
    8/1/1999 12:00:00 AM
  • Firstpage
    1165
  • Lastpage
    1171
  • Abstract
    Pinhole apertures made of depleted uranium were evaluated for high-resolution SPECT brain tumor imaging of therapeutic doses (30-100 mCi) of intratumorally administered I-131 radiolabeled monoclonal antibodies. High resolution is required for imaging residual tumor tissue, although for therapeutic doses, high collimator sensitivity is less important. Resolution of uranium and tungsten pinhole apertures was compared in planar and SPECT images of I-131 sources. Planar sensitivity was studied on- and off-axis. On-axis planar resolution for uranium pinholes is higher than that of tungsten pinholes for all cases examined; full-width-half-max (EWHM) is narrower for a 3 mm uranium pinhole than for a 1 mm tungsten pinhole, indicating significantly lower penetration at pinhole edges for uranium than for tungsten. Collimator sensitivity is higher for tungsten than for uranium, which also indicates higher edge penetration for tungsten. On-axis SPECT (14.4 cm radius) scans also show better resolution for a 2 mm uranium pinhole (7.8 mm EWHM) than for a 2 mm tungsten pinhole (9.5 mm FWHM), Background radiation from a uranium collimator (102 cts/sec) is acceptably low compared to expected clinical count rates
  • Keywords
    biomedical equipment; brain; single photon emission computed tomography; tumours; uranium; 14.4 cm; 2 to 9.5 mm; 3E-2 to 0.1 ci; I; I-131 SPECT imaging; U; W; collimator sensitivity; edge penetration; expected clinical count rates; full-width-half-max; intratumorally administered I-131 radiolabeled monoclonal antibodies; medical diagnostic imaging; nuclear medicine; off-axis; on-axis; planar images; planar sensitivity; residual tumor tissue imaging; therapeutic doses; uranium pinhole collimators; Apertures; Cancer; Collimators; High-resolution imaging; Image resolution; Lead; Neoplasms; Optical imaging; Spatial resolution; Tungsten;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.790851
  • Filename
    790851