DocumentCode :
1558709
Title :
Study of the resolution performance of an array of discrete detectors with independent readouts for positron emission tomography
Author :
Lecomte, R. ; Martel, C. ; Cadorette, J.
Author_Institution :
Dept. of Nucl. Med. & Radiol., Sherbrooke Univ., Que., Canada
Volume :
10
Issue :
3
fYear :
1991
fDate :
9/1/1991 12:00:00 AM
Firstpage :
347
Lastpage :
357
Abstract :
The intrinsic resolution characteristics of an original detection system for very high resolution positron emission tomography (PET) based on arrays of small discrete crystals individually coupled to solid state photodetectors are evaluated. The results obtained demonstrate the excellent three-dimensional (3-D) resolving capability of a detection system based on the EG&G/RCA C30994 detector module. The analysis of the various components contributing to the resolution has shown that the geometry of the detector and the intercrystal scattering both contribute to limit the achievable intrinsic resolution. The geometrical component is dominant off the central region of the camera field as a result of intercrystal penetration when the array is irradiated at an angle. Intercrystal scattering must be properly rejected to achieve the intrinsic resolution determined by the detector geometry
Keywords :
biomedical equipment; computerised tomography; radioisotope scanning and imaging; 3D resolution; detector geometry; discrete detectors array; intercrystal penetration; intercrystal scattering; medical diagnostic imaging; medical instrumentation; nuclear medicine; positron emission tomography; resolution performance; solid state photodetectors; Cameras; Degradation; Detectors; Energy resolution; Event detection; Image resolution; Light scattering; Positron emission tomography; Sensor arrays; Signal resolution;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.97584
Filename :
97584