DocumentCode :
1590082
Title :
Relative Location Estimation of Power-Receiver from Transmitter for In Vivo Robotic Capsules
Author :
Hwang, Ji-Su ; Kim, Jong-Dae ; Park, Chan-Young ; Ryu, Munho ; Kim, Jongwon
Author_Institution :
Div. of Inf. & Commun. Eng., Hallym Univ., Chunchon
fYear :
2006
Firstpage :
6635
Lastpage :
6637
Abstract :
This paper presents an estimation method of the relative location of the power-receiver from the transmitter for in vivo robotic capsules. The proposed method utilizes the fact that the needed current for driving the power transmitter is increased when the receiver coil is inductively coupled with the transmitter coil. By investigating the current increase characteristic according to the relative locations of the receiver, we draw an estimation strategy on the relative location of power-receiver from the transmission coil
Keywords :
endoscopes; medical robotics; receivers; transmitters; endoscopes; in vivo robotic capsules; power receiver; power transmitter; receiver coil; relative location estimation; transmitter coil; Coils; Couplings; Current measurement; Endoscopes; In vivo; Magnetic fields; Power amplifiers; Pulse width modulation; Robots; Transmitters;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location :
Shanghai
Print_ISBN :
0-7803-8741-4
Type :
conf
DOI :
10.1109/IEMBS.2005.1616023
Filename :
1616023
Link To Document :
بازگشت