DocumentCode :
161574
Title :
Wireless power transfer system applied to an active implantable medical device
Author :
Campi, Tommaso ; Cruciani, Silvano ; Feliziani, Mauro ; Hirata, Akihiko
Author_Institution :
Dept. of Ind. & Inf. Eng. & Econ., Univ. of L´Aquila, L´Aquila, Italy
fYear :
2014
fDate :
8-9 May 2014
Firstpage :
134
Lastpage :
137
Abstract :
This study deals with the design of a Wireless Power Transfer (WPT) system based on magnetic resonant coupling between two coils, whose secondary is located inside the human body and connected to a battery recharge system of an active implantable medical device (AIMD). A co-simulation tool is developed using circuit and electromagnetic field models to simulate the WPT system. Simulation results are obtained for different positions of the secondary coil and are compared with experiments.
Keywords :
coils; electromagnetic field theory; inductive power transmission; prosthetics; secondary cells; AIMD; WPT system; active implantable medical device; battery recharge system; circuit field models; cosimulation tool; electromagnetic field models; magnetic resonant coupling; secondary coil; wireless power transfer system; Biological tissues; Coils; Couplings; Integrated circuit modeling; Muscles; Skin; Wireless communication; Wireless power transfer (WPT); active implantable medical device (AIMD); battery recharging; magnetic field numerical solution;