Title :
Design and Implementation of a New Type Excitation Source and the Optimal Excitation Coil for MIT
Author :
Jiao, Mingke ; Qin, Mingxin ; Liang, Wenwen ; Zhang, Hua ; Li, Ke ; Zhou, Wei ; Yang, Hao ; Wang, Wenyong
Author_Institution :
Dept. of Biomed. Eng., Fourth Mil. Med. Univ., Xian
Abstract :
In this study, we have designed a high-performance excitation source and chose the optimal excitation coil to improve the efficiency of magnetic induction tomography system. A new type excitation source with adjustable output has been designed, the power circuit is simulated by the software ICPA. Focused and solenoid coils were designed by the design principle, then the optimal excitation coil was proved by measuring the magnetic field distribution and the experimentation of phase detection with cells models. At certain excitation frequency, parameters of output power of the excitation source are 0.035 W~31.4 W, the steady output peak current over 1 A, frequency stability of 10"9, THD amplitude of less -51dB. The focused discal excitation coil is most effective for phase detection with cells models than other coils. The excitation source can produce certain frequency sine wave, with higher frequency stability, lower THD and wider adjustable output power. The phase difference of normal and edema cell can be obtained by focused discal coil.
Keywords :
biomagnetism; biomedical measurement; cellular biophysics; neurophysiology; tomography; cells models; edema cell; focused discal coil; magnetic field distribution; magnetic induction tomography; optimal excitation coil; phase detection; solenoid coils; Circuit simulation; Coils; Frequency; Magnetic field measurement; Phase detection; Phase measurement; Power generation; Solenoids; Stability; Tomography;
Conference_Titel :
Bioinformatics and Biomedical Engineering, 2008. ICBBE 2008. The 2nd International Conference on
Conference_Location :
Shanghai
Print_ISBN :
978-1-4244-1747-6
Electronic_ISBN :
978-1-4244-1748-3
DOI :
10.1109/ICBBE.2008.919