DocumentCode :
1696146
Title :
Microfabricated iridium oxide reference electrode for continuous glucose monitoring sensor
Author :
Haesik Yang ; Sun Kil Kang ; Dong-Ho Shin ; Hyokyum Kim ; Youn Tae Kim
Author_Institution :
BioMEMS Team, Electron. & Telecommun. Res. Inst., Daejeon, South Korea
Volume :
1
fYear :
2003
Firstpage :
103
Abstract :
A novel microfabricated iridium oxide(IrO/sub x/) reference electrode is introduced which can be used in any buffer solutions. It is formed on Pt electrode by electrodeposition of IrO/sub x/ easily and uniformly. Its potential is stable in a phosphate buffered saline (PBS) solution and in serum after initial stabilization, and its potential performance is very reproducible. The behavior of a microfabricated glucose sensor employing an IrO/sub x/ reference electrode is similar to that employing a commercial Ag/AgCl reference electrode. All results shows that IrO/sub x/ could be used in continuous glucose monitoring sensors as a reference electrode.
Keywords :
biosensors; electrochemical electrodes; electrochemical sensors; electrochemistry; electrodeposits; iridium compounds; microelectrodes; pH; patient monitoring; platinum; proteins; IrO/sub x/; Pt; buffer solutions; electrodeposition; glucose monitoring sensor; microfabricated iridium oxide reference electrode; phosphate buffered saline; Amperometric sensors; Biosensors; Blood; Chemicals; Electrodes; Glass; In vivo; Monitoring; Stability; Sugar;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
TRANSDUCERS, Solid-State Sensors, Actuators and Microsystems, 12th International Conference on, 2003
Conference_Location :
Boston, MA, USA
Print_ISBN :
0-7803-7731-1
Type :
conf
DOI :
10.1109/SENSOR.2003.1215263
Filename :
1215263
Link To Document :
بازگشت