DocumentCode :
1755591
Title :
A Bi-Dimensional Multi-Wire Cathode Strip Detector for Fission Fragments
Author :
Vind, R.P. ; Biswas, D.C. ; Gupta, Y.K. ; Inkar, A.L. ; Jangale, R.V. ; Prajapati, G.K. ; Joshi, B.N. ; John, B.V. ; Nayak, B.K. ; Choudhury, R.K.
Author_Institution :
Nucl. Phys. Div., Bhabha Atomic Res. Centre, Mumbai, India
Volume :
60
Issue :
6
fYear :
2013
fDate :
Dec. 2013
Firstpage :
4650
Lastpage :
4655
Abstract :
A large area bi-dimensional position sensitive multi-wire cathode strip detector (MCSD) has been developed for the detection of fission fragments (FFs). The performance of the detector has been characterized with FFs from 252Cf and α particles from 241Am-239Pu sources. The position information is obtained from the measurement of time delay signals of the cathode strips (X-direction) and split-cathode wires (Y-direction) with respect to the timing signal from the anode wires. The position resolution is found to be about 1.0 and 1.5 mm in X-and Y-directions respectively.
Keywords :
alpha-particle detection; position sensitive particle detectors; radioactive sources; 239Pu source; 241Am source; 252Cf; alpha particle source; anode wire timing signal; fission fragment detection; large area bidimensional position sensitive multiwire cathode strip detector; position resolution; split-cathode wires; time delay signal measurement; Anodes; Cathodes; Delay lines; Gas detectors; Position measurement; Strips; Wires; Delay line read out method; fission fragment detection; multi-wire cathode strip detector; position sensitive gas detector;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2013.2282964
Filename :
6661414
Link To Document :
بازگشت