Abstract :
Needle steering in medical procedures has attracted considerable attention in recent years. For example, in prostate brachytherapy, it is desired to insert a flexible beveled-tip needle with minimum deflection. To date, different methods of insertion which incorporate needle rotation about its insertion axis have been proposed in order to reduce needle deflection and target displacement. In this paper, needle deflection resulting from different methods of insertion are compared with our ";model-based"; method which estimates the amount of needle deflection using Euler-Bernoulli beam equations. Experiments are performed in gelatin phantoms and animal tissue. The results show that the proposed ";model-based"; method reduces the amount of needle deflection more than other methods. In this paper, some factors for choosing the appropriate method of insertion are also discussed.
Keywords :
biological tissues; brachytherapy; phantoms; Euler-Bernoulli beam equations; animal tissue; flexible beveled-tip needle; gelatin phantoms; insertion methods; medical procedures; model-based method; needle deflection; needle rotation; prostate brachytherapy; target displacement; Biological tissues; Brachytherapy; Force measurement; Force sensors; Imaging phantoms; Magnetic sensors; Medical diagnostic imaging; Needles; Spinning; Surgery; Animals; Humans; Models, Theoretical; Monitoring, Intraoperative; Needles;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE
Conference_Location :
Lyon
ISSN :
1557-170X
Print_ISBN :
978-1-4244-0787-3
Type :
conf
DOI :
10.1109/IEMBS.2007.4352330
Filename :
4352330
Link To Document :
بازگشت