  • Abstract
    We evaluated image performance of scintillator coupled CMOS APS imager for radiography and mammography conditions, respectively. 80 kVp and 28 kVp of X-ray tube voltage at the SID of 720 mm and 600 mm were set for each condition. Commercially available scintillator such as Lanex screen, needle structured CsI(Tl), and fiber optic structured CsI (Tl) were coupled CMOS APS imager. From the experimental results, needle CsI type scintillator was candidate in radiography condition, having 45 % at 3 mm-1. However, at the higher frequency no scintillator showed good DQE in mammography condition although Lanex fine scintillator has DQE of 0.5 at 6 mm-1.
  • Keywords
    CMOS image sensors; diagnostic radiography; mammography; medical image processing; solid scintillation detectors; Lanex fine scintillator; SID; X-ray tube; detective quantum efficiency; digital radiography; fiber optic structured CsI (Tl); image performance; mammography; needle structured CsI(Tl); scintillator coupled CMOS APS imager; CMOS image sensors; Charge coupled devices; Image resolution; Mammography; Needles; Optical fibers; Pixel; Radiography; Sensor arrays; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352575
  • Filename
    1352575