• DocumentCode
    1844827
  • Title

    Development of a lead X-ray compensation method in simultaneous Tl-201 SPECT and Tc-99m TCT using a flood source

  • Author

    Motomura, N. ; Ichihara, T. ; Hasegawa, H. ; Ogawa, K. ; Hashimoti, J. ; Kubo, A.

  • Author_Institution
    Toshiba Med. Eng. lab., Tochigi, Japan
  • Volume
    2
  • fYear
    1996
  • fDate
    2-9 Nov 1996
  • Firstpage
    1416
  • Abstract
    We have developed a compensation method for contamination in Tl-201 data due to lead X-ray in simultaneous Tl-201 SPECT and Tc-99m TCT using a flood source. The evaluation of the proposed method was performed with a chest phantom and a normal volunteer. The significant underestimation in Tl-201 SPECT data due to the lead X-ray could be improved with the proposed method. The simultaneous Tl-201 SPECT images compensated using the proposed method were almost the same as the Tl-201 alone images. As a result, the feasibility of the proposed method was demonstrated
  • Keywords
    computerised tomography; lead; single photon emission computed tomography; technetium; thallium; 201Tl SPECT; 99Tcm TCT; Tc; Tl; lead X-ray compensation method; simultaneous single photon emission computed tomography; transmission computed tomography; Attenuation; Biomedical engineering; Collimators; Contamination; Electromagnetic scattering; Floods; Gamma rays; Particle scattering; Single photon emission computed tomography; X-ray scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-3534-1
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1996.591713
  • Filename
    591713