• DocumentCode
    1880352
  • Title

    Long-term stability of EMG recordings using a wireless EMG electrode

  • Author

    Skov-Madsen, Gitte ; Rijkhoff, Nico ; Vistisen, Bodil K.

  • Author_Institution
    Center for Sensory-Motor Interaction (SMI), Aalborg Univ., Aalborg
  • fYear
    2008
  • fDate
    25-28 Oct. 2008
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    The stability of EMG recordings over an 8-day period has been investigated using a new wireless EMG electrode in 10 healthy volunteers. Throughout the 8-day period the electrode stayed on the skin, only the battery was replaced before each recording session. EMG-RMS values were recorded from the biceps during 50% and 100%-MVC exercises on 6 out of the 8 days. The results show a gradual decrease in RMS values with time. The RMS values at day 8 were, compared with day 1, down with 27.1% and 31.7% for 50% MVC and 100% MVC, respectively.
  • Keywords
    biomedical electrodes; electromyography; sensors; EMG recordings; EMG sensor; biceps; exercises; long-term stability; time 8 day; wireless EMG electrode; Batteries; Electrodes; Electromyography; Muscles; Reproducibility of results; Silver; Skin; Stability; Testing; Wireless sensor networks; EMG; Long-term; biceps; surface;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Applied Sciences on Biomedical and Communication Technologies, 2008. ISABEL '08. First International Symposium on
  • Conference_Location
    Aalborg
  • Print_ISBN
    978-1-4244-2647-8
  • Electronic_ISBN
    978-1-4244-2648-5
  • Type

    conf

  • DOI
    10.1109/ISABEL.2008.4712606
  • Filename
    4712606