DocumentCode :
1884961
Title :
Self-Calibration Methods in Measurement Channels for Linear and Non-Linear Conversion Functions Case
Author :
Gawedzki, Waclaw
Author_Institution :
Dept. of Meas. & Instrum., Univ. of Sci. & Technol., Krakow
fYear :
2006
fDate :
24-27 April 2006
Firstpage :
1506
Lastpage :
1510
Abstract :
Self-calibration methods for long-term monitoring of non-electrical quantities are presented in the paper. There have been discussed the self-calibration methods of measurement channels containing integrated ratiometric output sensors and ratiometric output sensors with strong attenuation (bridge circuits). It has been shown that under certain system solutions this type sensors make measurement results independent from possible changes of the parameters of elements of electronic measuring systems caused by the temporal and thermal instability, and, also, the measurement accuracy improves. In the paper an algorithm is proposed, which enables self-calibration of measurement channels with non-linear conversion functions
Keywords :
calibration; measurement systems; sensors; bridge circuits; electronic measuring systems; integrated ratiometric output sensors; linear conversion functions; measurement channels; nonelectrical quantities; nonlinear conversion functions; self-calibration methods; thermal instability; Bridge circuits; Capacitive sensors; Computerized monitoring; Frequency conversion; Frequency measurement; Integrated circuit measurements; Pollution measurement; Sensor systems; Strain measurement; Voltage; non-electrical monitoring; self-calibration; sensors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2006. IMTC 2006. Proceedings of the IEEE
Conference_Location :
Sorrento
ISSN :
1091-5281
Print_ISBN :
0-7803-9359-7
Electronic_ISBN :
1091-5281