DocumentCode
1899640
Title
Laboratory panel and radiometer calibration
Author
Deadman, A.J. ; Behnert, I.D. ; Fox, N.P. ; Griffith, D.
Author_Institution
Nat. Phys. Lab. (NPL), Teddington, UK
fYear
2011
fDate
24-29 July 2011
Firstpage
3883
Lastpage
3886
Abstract
This paper presents the results of the laboratory and field based reflectance panel and radiometer comparisons that took place as part of the CEOS 2010 Key comparison of "techniques and instruments used for the vicarious calibration of Land surface imaging through a ground reference standard test site". The results of the comparisons are presented which shows that the different ways in which reflectance panels are calibrated can give different results for the reflectance attributed to a test site and that changing illumination and environmental conditions can effect the measured target reflectance.
Keywords
calibration; radiometers; reflectivity; remote sensing; topography (Earth); CEOS 2010; field based reflectance panel; ground reference standard test site; laboratory panel; land surface imaging; radiometer calibration; target reflectance; Atmospheric measurements; Calibration; Laboratories; Lighting; Radiometry; Reflectivity; Sun; Earth Observation; Field spectroscopy; Reflectance factor; Vicarious calibration; goniometric;
fLanguage
English
Publisher
ieee
Conference_Titel
Geoscience and Remote Sensing Symposium (IGARSS), 2011 IEEE International
Conference_Location
Vancouver, BC
ISSN
2153-6996
Print_ISBN
978-1-4577-1003-2
Type
conf
DOI
10.1109/IGARSS.2011.6050079
Filename
6050079
Link To Document