Application of the Monte Carlo method to estimate the tenth-value thickness for X-rays in medical electron accelerators
Author
Ródenas, José ; Verdú, Gumersindo
Author_Institution
Dept. de Ingenieria Quimica y Nucl., Valencia Univ., Spain
fYear
1993
fDate
17-20 May 1993
Firstpage
647
Abstract
DIN-6847 is the most suitable standard to estimate shielding requirements in medical radiotherapy installations using linear electron accelerators. Its calculation method is based on the tenth-value thickness concept for which values given by curves and tables are recommended in DIN-6847. These parameters have been estimated for both primary and secondary X-ray beams using the MCNP code based on the Monte Carlo method. Results show that DIN-6847 values are conservative for secondary X-ray beam, but not always for direct radiation. The methodology used has been tested using the mentioned code and results are consistent
Keywords
Monte Carlo methods; electron accelerators; linear accelerators; radiation therapy; DIN-6847; Monte Carlo method; X-rays; linear electron accelerators; medical electron accelerators; medical radiotherapy installations; primary X-ray beams; secondary X-ray beams; shielding requirements; tenth-value thickness; Aluminum; Electromagnetic scattering; Electron accelerators; Electron beams; Linear accelerators; Linear particle accelerator; Parameter estimation; Particle scattering; Testing; X-ray scattering;
fLanguage
English
Publisher
ieee
Conference_Titel
Particle Accelerator Conference, 1993., Proceedings of the 1993
Conference_Location
Washington, DC
Print_ISBN
0-7803-1203-1
Type
conf
DOI
10.1109/PAC.1993.308828
Filename
308828
Link To Document