DocumentCode
1978203
Title
The Fringing Capacitance of Reference Open Circuits Used in Network Analyser Calibration
Author
Ambrose, C. ; Aitchison, C.S.
Author_Institution
Department of Electronics, Chelsea College, University of London, Pulton Place, London SW6 5PR, ENGLAND
fYear
1980
fDate
8-12 Sept. 1980
Firstpage
273
Lastpage
277
Abstract
The effect of incorrectly calibrating an automatic network analyser by using a wrong value for the open-circuit fringing capacitance has been studied. It was found that any measurement made using this calibration will be in error by an amount that is independent of the network analyser error network. A new linear frequency dependent value for the fringing capacitance of an APC-7 shielded open-circuit has been established.
Keywords
Calibration; Capacitance measurement; Circuits; Differential equations; Error analysis; Error correction; Frequency measurement; Impedance measurement; Intelligent networks; Reflection;