• DocumentCode
    1995102
  • Title

    Reproducibility of impedance cardiogram waveform analysis in lower-body and upper-body exercise

  • Author

    Kizakevich, Paul N. ; Hazucha, Milan ; Van Hoose, Linda ; Bolick, Karen ; Jochem, Warren J. ; McCartney, Michael ; McMaster, Lany

  • Author_Institution
    Res. Triangle Inst., Research Triangle Park, NC, USA
  • fYear
    1994
  • fDate
    10-12 Jun 1994
  • Firstpage
    182
  • Lastpage
    187
  • Abstract
    Measurements of nine impedance cardiograph variables made on 17 young (21-29 yr, mean 24.6 yr), healthy males during rest and three levels of treadmill and hand-crank exercise showed no significant difference (ANOVA, p>0.05) among the three days, except for heart rate (p<0.01, treadmill; p<0.02, hand crank), time-to-aortic closing (p<0.01, treadmill), and mean thoracic impedance (p<0.01, hand crank). Stroke volume, cardiac output, time-to-peak ejection, time-to-aortic opening, (dZ/dt)peak, and d2Z/dt 2 were invariant with day of exercise
  • Keywords
    biomedical measurement; cardiology; electric impedance measurement; medical signal processing; waveform analysis; 21 to 29 y; cardiac output; hand-crank exercise; heart rate; impedance cardiogram waveform analysis reproducibility; lower-body exercise; mean thoracic impedance; stroke volume; time-to-aortic closing; time-to-aortic opening; time-to-peak ejection; treadmill exercise; upper-body exercise; young healthy males; Analysis of variance; Cardiography; Cardiology; Costs; Electrocardiography; Heart rate; Impedance measurement; Protocols; Reproducibility of results; Signal processing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computer-Based Medical Systems, 1994., Proceedings 1994 IEEE Seventh Symposium on
  • Conference_Location
    Winston-Salem, NC
  • Print_ISBN
    0-8186-6256-5
  • Type

    conf

  • DOI
    10.1109/CBMS.1994.316008
  • Filename
    316008