• DocumentCode
    2001987
  • Title

    Combined fan/parallel beam reconstruction in cardiac SPECT imaging

  • Author

    Narayanan, M.V. ; King, Michael A. ; Byrne, Charles L.

  • Author_Institution
    Massachusetts Univ. Med. Center, Worcester, MA, USA
  • Volume
    3
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    1142
  • Abstract
    With three-headed SPECT systems, two heads can aquire truncation-free emission images with parallel-beam (PB) collimators, while the third head acquires transmission images through a fan-beam (FB) collimator. The FB collimator can also be used for emission imaging. Combining FB and PB emission data from the 3 heads, offers the potential for improved sensitivity with the trade-off being truncation artifacts. In this work, the authors investigate 3 methods to combine fan and parallel beam (F&PB) reconstructions. The first approach uses a PB estimate as a prior for the regularized AB-EMML algorithm to reduce truncation in the FB reconstruction. The second approach combines the parallel and re-sorted FB projections and reconstructs a composite image. The third method is based on the work of Jasczak el al. (Phys. Med. Biol., vol. 37, p. 535-48, 1992) where the intermediate F&PB reconstructions are combined to reduce the truncation artifacts inherent In FB imaging. MCAT simulation results indicate that the combined F&PB reconstructions provide decreased image noise while greatly limiting the truncation artifacts inherent in FB imaging
  • Keywords
    cardiology; image reconstruction; medical image processing; single photon emission computed tomography; cardiac SPECT imaging; combined fan/parallel beam reconstruction; composite image; decreased image noise; improved sensitivity; medical diagnostic imaging; nuclear medicine; three-headed SPECT systems; transmission images; truncation artifacts; truncation-free emission images; Attenuation; Collimators; Geometry; Image reconstruction; Imaging phantoms; Maximum likelihood detection; Maximum likelihood estimation; Noise level; Torso; Upper bound;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.842762
  • Filename
    842762