Title :
Mean arterial pressure estimation method using morphological changes in oscillometric waveform
Author :
Song, SH ; Kim, DK ; Lee, JS ; Chee, YJ ; Kim, IY
Author_Institution :
Dept. of Biomed. Eng., Hanyang Univ., Seoul, South Korea
Abstract :
Oscillometric method is the method for the blood pressure measurement using mean arterial pressure (MAP). In the conventional method, the cuff pressure at the largest pulse is determined as MAP. However, it is difficult to find the largest oscillometric pulse from patients who have the cardiovascular disease or the elder subjects. Therefore, we studied a new method to determine the MAP by analyzing the changes of the morphology of oscillometric pulse different from above method. The new method was applied to the cuff pressure recorded by 11 healthy subjects. Two observers trained by BHS program measured the blood pressure using the auscultatory method to determine the reference MAP. The mean difference and standard deviation of the MAP determined by the algorithm proposed by this study were 1.45 mmHg and 3.47 mmHg respectively.
Keywords :
blood pressure measurement; blood vessels; diseases; blood pressure measurement; cardiovascular disease; elder subjects; mean arterial pressure estimation; oscillometric pulse morphology changes; oscillometric waveform; Biomedical engineering; Biomedical measurements; Blood pressure; Cardiovascular diseases; Equations; Hypertension; Morphology; Particle measurements; Pressure measurement; Space vector pulse width modulation;
Conference_Titel :
Computers in Cardiology, 2009
Conference_Location :
Park City, UT
Print_ISBN :
978-1-4244-7281-9
Electronic_ISBN :
0276-6547