• DocumentCode
    2052019
  • Title

    An improved PI-method for reconstruction from helical cone-beam projections

  • Author

    Turbell, H. ; Danielsson, P.E.

  • Author_Institution
    Image Process. Lab., Linkoping Univ., Sweden
  • Volume
    2
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    865
  • Abstract
    Aiming for a more coherent 1D-filtering, the authors propose an improvement of the original PI-method. A PI-surface is defined by the object points which enter and exit over detector edges simultaneously. In between these to events the PI-surface is projected onto the detector, not quite along a perfect line, but nevertheless with most of the points concentrated along a certain non-horizontal line. The authors show that one-dimensional filtering along these slanted lines improves the image quality considerably
  • Keywords
    computerised tomography; image reconstruction; medical image processing; CT; PI-method; detector edges; image quality improvement; medical diagnostic imaging; nonhorizontal line; object points; one-dimensional filtering; slanted lines; Detectors; Event detection; Filtering; Geometry; Image edge detection; Image processing; Image quality; Image reconstruction; Laboratories; Object detection;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.845801
  • Filename
    845801