Abstract :
The Finnish Centre for Metrology and Accreditation (MIKES) has two HP 5071A Cs-beam standards (standard version from 1997 and high stability version from 2000) and one CH1-76 Passive hydrogen maser (since February 2003). A special four-channel phase comparator with better than 0.3 ps single shot resolution is used to study the short-term stability of the Cs-standards, with hydrogen maser as a reference. Long term stability is measured with GPS as a reference. Allan stability plots indicate that the high stability version 5071A is approximately three times better than the standard version. The high stability version has only slightly smaller long term drift (ageing) than standard version but frequency retrace after clock failure or power blackout is significantly better in high stability version.
Keywords :
atomic clocks; masers; measurement standards; phase comparators; stability; Allan stability; CH1-76 Passive hydrogen maser; Finnish Centre for Metrology and Accreditation; GPS; HP 5071A caesium-beam standards; Long term stability; clock failure; environmental effects; four-channel phase comparator; frequency retrace; power blackout; short-term stability; Allan stability; Cesium clock; Environmental effects;
fLanguage :
English
Publisher :
iet
Conference_Titel :
Frequency and Time Forum, 2004. EFTF 2004. 18th European
Conference_Location :
Guildford
ISSN :
0537-9989
Print_ISBN :
0-86341-384-6
Type :
conf
Filename :
5074984
Link To Document :
بازگشت