Abstract :
Piezoelectric transducers used in therapy, for example in lithotripsy, consist of several thousand cylindrical piezoceramic elements which are driven by a high voltage pulse. In order to simulate the radiated pressure signal at the transducer surface the transient behaviour of a single cylindrical piezoceramic element was investigated using finite element analysis (FEA). The realistic electrical load conditions are integrated as well as the mechanical load given by the construction. Results show that the element does not behave like a piston source. Measurements agree to simulation results very well. It is concluded that the simulation results could provide valuable information for further design improvements of therapeutic transducers
Keywords :
biomedical transducers; biomedical ultrasonics; finite element analysis; piezoceramics; piezoelectric transducers; radiation therapy; transient response; ultrasonic transducers; cylindrical piezoceramics; lithotripsy; mechanical load; realistic electrical load conditions; self focusing transducers; single cylindrical piezoceramic element; therapeutic transducer design improvement; therapy transducers; Analytical models; Finite element methods; Lithotripsy; Medical treatment; Piezoelectric materials; Piezoelectric transducers; Signal analysis; Transient analysis; Transient response; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 1999. Proceedings. 1999 IEEE
Conference_Location :
Caesars Tahoe, NV
ISSN :
1051-0117
Print_ISBN :
0-7803-5722-1
Type :
conf
DOI :
10.1109/ULTSYM.1999.849222
Filename :
849222
Link To Document :
بازگشت