Title :
A head motion measurement system suitable for 3D cone-beam tomography using markers
Author :
Bhowmik, Ujjal Kumar ; Adhami, Reza R.
Author_Institution :
Dept. of Electr. & Comput. Eng., Univ. of Alabama in Huntsville, Huntsville, AL, USA
fDate :
Aug. 28 2012-Sept. 1 2012
Abstract :
Head motion during brain CT studies can degrade the reconstructed image through distortion and other artifacts such as blurring, doubling and thereby contributing to misdiagnosis of diseases. Estimation of motion parameters is essential for mitigating motion artifacts. In this paper, we propose a marker based numerical optimization method to measure six degrees of freedom of head motion in three-dimensional cone-beam CT system without using any external motion sensors. Simulation results demonstrate that our method has prerequisite accuracy, linearity and range compared to the existing external sensor based method.
Keywords :
biomechanics; computerised tomography; medical image processing; optimisation; external motion sensor; external sensor based method; head motion measurement system; marker based numerical optimization method; three-dimensional cone-beam tomography; Accuracy; Approximation methods; Computed tomography; Detectors; Head; Linearity; Simulation; Head Movements; Humans; Tomography, X-Ray Computed;
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Conference_Location :
San Diego, CA
Print_ISBN :
978-1-4244-4119-8
Electronic_ISBN :
1557-170X
DOI :
10.1109/EMBC.2012.6347355