Title :
Communication and storage of compressed resting and exercise ECGs using the revised SCP-ECG standard
Author :
Zywietz, C. ; Fischer, R. ; Joseph, G.
Author_Institution :
Biosignal Process., Med. Sch., Hannover, Germany
Abstract :
Exercise ECG (XECG) testing is one of the clinical routine examinations which produces a large amount of data if full-disclosure records aree to be provided. We describe a system for XECG analysis which takes advantage of the data compression procedures and the data format of the SCP-ECG (Standard Communication Protocol for computer-assisted Electrocardiography) protocol. Processing the XECG data in consecutive 10-s data sections resulted in six SCP blocks per minute. Over a test set of 20 XECGs with various test durations, a median compression ratio of 11.1 has been achieved. We have also investigated the influence of noise on the compression ratio. 50 μV RMS may reduce the compression ratio from ~17 to ~8
Keywords :
PACS; electrocardiography; medical signal processing; source coding; telecommunication standards; transport protocols; 10 s; clinical routine examinations; compressed ECG communication; compressed ECG storage; compression ratio; data compression procedures; data format; data sections; exercise ECG testing; full disclosure records; noise; resting ECG; revised SCP-ECG standard; test durations; Communication standards; Data compression; Electrocardiography; Heart rate; Medical tests; Noise reduction; Protocols; Rhythm; Signal to noise ratio; Testing;
Conference_Titel :
Computers in Cardiology 2001
Conference_Location :
Rotterdam
Print_ISBN :
0-7803-7266-2
DOI :
10.1109/CIC.2001.977623