Abstract
Comparisons of the 1 V Josephson array voltage standard (JAVS) of the "D.I. Mendeleyev Institute for Metrology" (VNIIM), Russia, with that of the Physikalisch-Technische Bundesanstalt (PTB), Germany, were carried out in St. Petersburg in June 1999. The results of the direct comparison agree within an overall standard uncertainty of less than 1 nV. The results also show that direct comparisons are useful for detecting voltage offsets of a few parts in 10/sup 10/.
Keywords
Josephson effect; Zener diodes; measurement standards; measurement uncertainty; voltage measurement; 1 nV; Josephson voltage standards; Mendeleyev Institute for Metrology; PTB; Physikalisch-Technische Bundesanstalt; VNIIM; Zener reference; direct comparison; indirect comparison; standard uncertainty; voltage offsets; Detectors; Frequency; Lead; Metrology; Oscillators; Signal generators; Switches; Thermal resistance; Uncertainty; Voltage;